Umweltanalytik Holbach GmbH

Technical guide

Counting Mould Spores

  • MBASS30V3 with PS 30 particle sampling head

To count airborne mould spores by microscopy, collect particles on an adhesive microscope slide using the PS 30 head on the MBASS30V3. The laboratory examines the deposit directly. Spores can therefore be included even when they would not form a colony on the chosen culture medium.

At a glance

Application Airborne mould spores: sampling for a total spore count
Equipment MBASS30V3, PS 30, coated slides
Laboratory assessment Direct microscopy

Choose the method before the sampler

A culture result describes colonies recovered on a specified medium under specified conditions. A microscopic count describes particles that an analyst can recognise in the deposit. The two results answer different questions. When comparing rooms or investigating a change over time, agree on one method and record the conditions that could affect the comparison.

Match the slide to the laboratory

Ask the analyst which slide, labelling system and submission information the laboratory accepts. The PS 30 can place up to three separate sampling traces on a compatible slide. Assign each trace its own location, sample volume and identifier before sending it for analysis. An unidentified trace cannot be assigned reliably to a room.

Read the report in context

The measured air volume lets the laboratory relate the count to the sampled volume. Microscopic appearance limits the level of identification. Read the result alongside the inspection, potential moisture sources and any suitable reference sample. DIN ISO 16000-20 describes sampling for the determination of the total spore count. Follow the specification applicable to the commission.

Can I use my existing MBASS30V3?

Yes. Ask for the PS 30 particle sampling head and suitable coated slides. Specify the head and slide holder already available to avoid duplicate equipment.

Choose a PS 30 for airborne mould spore sampling. Tell us your current sampler and which laboratory will examine the slides.

Sources and equipment

PS 30 · DIN ISO 16000-20 · Holbach technical documents